STATIC STRENGTH ESTIMATE OF STRUCTURAL ELEMENTS OF IMPLANT SYSTEMS LIKO 4 × 10 OF VARIOUS DESIGNS

https://doi.org/10.33667/2782-4101-2023-1-10-14

Аннотация

Introduction. The creation of effective, reliable, safe, technologically advanced and competitive products is the main task of medical device engineering. The most important requirement to modern medical devices is to guarantee patients’ safety during their lifetime. Today the use of modern computer-aided engineering analysis packages is the most effective calculation method for evaluating the strength and reliability of unique medical devices that can lead to serious consequences if their operation is disturbed. One of the most suitable and efficient systems for computer-aided engineering (CAE) system is the ANSYS software.
The purpose of this study was the comparative assessment of the elastic and elastoplastic formation of 4x10 dental implants of different designs on the abutment-pin and screw-body interface, using the computer simulation of the stress-strain state.
Materials and methods. Two kinds of dental implants were chosen for this study: a Liko-M 4 × 10 implant with the cylindrical body shape and a Liko-M DG 4 × 10 implant with a tapered body shape. The contact between the abutment and screw as well as the implant body and screw is frictional. The pre-tensioning of the screw from the initial tightening was 400 N. The load was applied to the cylindrical surface of the abutment at a percentage of its height.
Results. Elastic and elasto-plastic calculations of the stress-strain state of Liko-M 4 × 10 and Liko-M DG 4 × 10 implants were performed. Besides the results of the main calculations of the stress-strain state of the implants Liko-M 4x10 and Liko-M DG 4 × 10, necessary to assess their static strength, we have also calculated the strength coefficients of implant bodies. Comparative analysis of the static strength of the Liko-M 4 × 10 and Liko-M DG 4 × 10 implants provides conclusions, which are significant for practical application of the implants.

Об авторах

A. Sarkisov
Astrakhan State Medical University of the Ministry of Health of the Russian Federation
Россия


V. Avanisyan
Stavropol State Medical University of the Ministry of Health of the Russian Federation
Россия


A. Dolgalev
Stavropol State Medical University of the Ministry of Health of the Russian Federation
Россия


Yu. Sergeev
Stavropol State Medical University of the Ministry of Health of the Russian Federation
Россия


Список литературы

1. Pakharkov G. N., Popechitelev E. P. Principles and methods of quality assurance of medical and technical equipment of healthcare: Textbook. – St. Petersburg: Publishing house of St. Petersburg State Electrotechnical University «LETI», 2003.

2. Theory Reference for ANSYS and Workbench. – Canonsburg: ANSYS Inc.

3. Structural Analysis Guide. – Canonsburg: ANSYS Inc, 2019.

4. Bragin E. A., El’kanov A. A., Dolgalev A. A., Sergeev Y. A., Avanisyan V. M. Comparative assessment of static strength of implant-abutment connections of various implant shapes // Actual problems in dentistry. 2023. no. 1. pp. 121–125. DOI: https://doi.org/10.18481/2077–7566–2023–19–1–121–125


Рецензия

Для цитирования:


 ,  ,  ,   . International journal of Innovative Medicine. 2023;(1):10-14. https://doi.org/10.33667/2782-4101-2023-1-10-14

For citation:


Sarkisov A.A., Avanisyan V.M., Dolgalev A.A., Sergeev Yu.A. STATIC STRENGTH ESTIMATE OF STRUCTURAL ELEMENTS OF IMPLANT SYSTEMS LIKO 4 × 10 OF VARIOUS DESIGNS. International journal of Innovative Medicine. 2023;(1):10-14. https://doi.org/10.33667/2782-4101-2023-1-10-14

Просмотров: 165


Creative Commons License
Контент доступен под лицензией Creative Commons Attribution 4.0 License.


ISSN 2782-4101 (Online)